Insulation resistance meter MIC-2501.
Insulation resistance measurement:
- selectable measurement voltage 100. ..2500 V selected in steps of 100 V,
- continuous indication of the measured insulation resistance or leakage current,
- automatic discharge of the capacitance of the measured object after completing the insulation resistance measurement,
- acoustic determination of five-second time intervals to facilitate the capture of time characteristics,
- measured measurement times T1, T2 and T3 for the measurement of two absorption coefficients for 15, 60 and 600 s,
- indication of the actual measurement voltage during the measurement,
- protection against the measurement of live objects.
Insulation resistance measurement using the two- and three-wire method.
Low-voltage measurement of circuit continuity and resistance :
Leakage current measurement during insulation resistance measurement.
Measurement of direct and alternating voltages in the range of 0...750 V.
Memory of 990 cells (11880 entries) data transfer to a PC via USB cable .
Rechargeable battery.
Electrical safety:
- type of insulation: double, in accordance with PN-EN 61010 -1 and IEC 61557
- measurement category: IV 600 V (III 1000 V)
- degree of protection: IP65
Others technical data:
- meter power supply: SONEL L-1 NiMH 9.6 V battery pack; external power supply 12 V 2.5 A
- display: segment LCD
- memory of measurement results: 990 cells, 11880 entries,
- transmission of results: isolated USB connection< /li>